The compact high-sensitivity spectrometer WS400 is developed for low-light applications such as measurements of Raman scattering, fluorescence, etc., that require especially high sensitivity or enhanced specifications in the UV spectral range. 


Key Features:

  • Up-to-date model based on back-thinned arrays for operation with low signals

  • Enhanced UV sensitivity

  • Monolithic housing for optimum stability

  • Internal light traps to reduce stray light

  • Flexible configuration, including a choice of Diffraction gratings, spectral ranges and resolutions tailored to your requirement

  • User-friendly interface compatible with Windows XP, Windows 7, 8 and 10

Ws400 TECHNICAL FEATURES:

Detection & Image Sensors

Spectral resolution: depends on selected diffraction grating (see table below)

Image Sensors model S10420-1106-01
Hamamatsu
S16010-1106
Hamamatsu
S7030-1006
Hamamatsu
Number of pixels 2068 × 70 1024 × 64
Number of active pixels 2048 × 64 1024 × 58
Pixel size, μm 14 × 14 24 × 24
Active area size, mm 28.672 × 0.896 24.6 × 1.4
Maximum spectral sensitivity, nm 500 700 650
Maximum quantum efficiency, % >75 >90
Non-uniformity of sensitivity 1), % ±3 ±3
Anti-blooming 2) Yes No
ADC Resolution 16-bit, 250 kHz 16-bit, 125 kHz
Mean-square reading noise, ADC counts <4 <2
Dynamic range (binning mode) ~16000:1 ~33000:1
Exposure time, s 0.0083–10 3) 0.0082–2 3)
Frame processing time in the binning mode, ms 9.39 9.12
Thermoelectric cooling No
Operating temperature 10–30 °С

1) Signal level ≥ 50% of saturation.
2) Anti-blooming – sensor’s feature eliminating overflow of charges from over-exposed pixels to the neighboring ones.
3) Maximum storage time is deemed to be the time at which dark signal reaches 25% of the dynamic range at the ambient temperature +25°С.

Diffraction Grating

Diffraction grating, grooves/mm 1800 1200 900 600 400 300 200
Spectral range of grating operation, nm 190 – 600 190 – 900 190 – 1100 190 – 1100 190 – 1100 190 – 1100 190 – 1100
Multichannel array bandpass (average), nm 105 2)
92 3)
170 2)
145 3)
230 2)
200 3)
350 2)
300 3)
540 2)
460 3)
720 2)
620 3)
1000 2)
920 3)
Reciprocal linear dispersion (average), nm/mm 3.8 6.0 8.2 12.5 19.0 25.4 38.0
Spectral resolution (average), nm 1) 0.16 2)
0.22 3)
0.25 2)
0.36 3)
0.34 2)
0.5 3)
0.5 2)
0.75 3)
0.8 2)
1.1 3)
1.0 2)
1.5 3)
1.6 2)
2.3 3)

1) For 300nm wavelength.
2) For Spectrometer with S10420-1106.01 and S16010-1106 image sensor models.
3) For Spectrometer with the S7030-1006 image sensor model.