WS400
High-Sensitivity Compact Spectrometer
The compact high-sensitivity spectrometer WS400 is developed for low-light applications such as measurements of Raman scattering, fluorescence, etc., that require especially high sensitivity or enhanced specifications in the UV spectral range.
Key Features:
Up-to-date model based on back-thinned arrays for operation with low signals
Enhanced UV sensitivity
Monolithic housing for optimum stability
Internal light traps to reduce stray light
Flexible configuration, including a choice of Diffraction gratings, spectral ranges and resolutions tailored to your requirement
User-friendly interface compatible with Windows XP, Windows 7, 8 and 10
-
The high-aperture optical bench of the WS400 has no moving parts and minimizes the number of reflective surfaces. The WS400 features a monolithic housing, thus ensuring long-time measurement stability. The case has internal fins (light traps) that allow reducing stray light – a typical problem of compact short-focal length instruments.
The WS400 may contain as a detector either one of two non-cooled Hamamatsu back-thinned CCD-arrays without a charge-storage section (back-thinned FFT CCD), operating in the binning mode. Both arrays have wide spectral range and excellent sensitivity in the UV (see below for the curves of each detector spectral sensitivity and their specifications). These two arrays differ in quantum efficiency, data reading rate, dynamic range, and in pixel active area size.
-
Upon request, the WS400 spectrometer can be optionally calibrated for spectral sensitivity.
Available calibration ranges:
Cal-UV: 220–400
Cal-VIS: 300–1050 nm
Calibration is carried out using certified light sources—a deuterium lamp for the UV range and a halogen lamp for the VIS range. It is performed via fiber input and remains valid for one year, provided the optical fiber remains connected. Calibration accuracy is ±10%.
Ws400 TECHNICAL FEATURES:
-
Spectral range, nm: 190 — 1100
-
Focal length, mm: 125
F/number: 1 : 3,9
Entrance slit (one to choose): fixed width: 14μm, 30μm and 50 μm
Detection & Image Sensors
Spectral resolution: depends on selected diffraction grating (see table below)
| Image Sensors model | S10420-1106-01 Hamamatsu |
S16010-1106 Hamamatsu |
S7030-1006 Hamamatsu |
| Number of pixels | 2068 × 70 | 1024 × 64 | |
| Number of active pixels | 2048 × 64 | 1024 × 58 | |
| Pixel size, μm | 14 × 14 | 24 × 24 | |
| Active area size, mm | 28.672 × 0.896 | 24.6 × 1.4 | |
| Maximum spectral sensitivity, nm | 500 | 700 | 650 |
| Maximum quantum efficiency, % | >75 | >90 | |
| Non-uniformity of sensitivity 1), % | ±3 | ±3 | |
| Anti-blooming 2) | Yes | No | |
| ADC Resolution | 16-bit, 250 kHz | 16-bit, 125 kHz | |
| Mean-square reading noise, ADC counts | <4 | <2 | |
| Dynamic range (binning mode) | ~16000:1 | ~33000:1 | |
| Exposure time, s | 0.0083–10 3) | 0.0082–2 3) | |
| Frame processing time in the binning mode, ms | 9.39 | 9.12 | |
| Thermoelectric cooling | No | ||
| Operating temperature | 10–30 °С | ||
1) Signal level ≥ 50% of saturation.
2) Anti-blooming – sensor’s feature eliminating overflow of charges from over-exposed pixels to the neighboring ones.
3) Maximum storage time is deemed to be the time at which dark signal reaches 25% of the dynamic range at the ambient temperature +25°С.
Diffraction Grating
| Diffraction grating, grooves/mm | 1800 | 1200 | 900 | 600 | 400 | 300 | 200 |
| Spectral range of grating operation, nm | 190 – 600 | 190 – 900 | 190 – 1100 | 190 – 1100 | 190 – 1100 | 190 – 1100 | 190 – 1100 |
| Multichannel array bandpass (average), nm | 105 2) 92 3) |
170 2) 145 3) |
230 2) 200 3) |
350 2) 300 3) |
540 2) 460 3) |
720 2) 620 3) |
1000 2) 920 3) |
| Reciprocal linear dispersion (average), nm/mm | 3.8 | 6.0 | 8.2 | 12.5 | 19.0 | 25.4 | 38.0 |
| Spectral resolution (average), nm 1) | 0.16 2) 0.22 3) |
0.25 2) 0.36 3) |
0.34 2) 0.5 3) |
0.5 2) 0.75 3) |
0.8 2) 1.1 3) |
1.0 2) 1.5 3) |
1.6 2) 2.3 3) |
1) For 300nm wavelength.
2) For Spectrometer with S10420-1106.01 and S16010-1106 image sensor models.
3) For Spectrometer with the S7030-1006 image sensor model.
-
Triggering: internal / external
-
PC connection: Full Speed USB
Optical input:
— Direct radiation input to the entrance slit
— Optical Fiber : 0.6 (0.4)mm diameter, 1m length, SMA-905 connector ( As an option)
-
Dimensions, mm: 158 × 182 × 74
Weight, kg: 2.2
Similar Products:
-

WS100
-

WS200-IR
-

WS300
-

WS600