WM300

Automated monochromator-spectrograph

The WM300 monochromator is a versatile, high-aperture instrument offering excellent line quality and low stray light, with a 284 mm focal length and a 30×10 mm flat field. 


Key Features:

  • High aperture ratio (F=1:3.8)

  • Two output ports

  • Fully automated

  • The Wavelight.lab software

  • SDK, DLL

Wm300 TECHNICAL FEATURES:

Diffraction Gratings

  • Diffraction gratings: 50x50x10mm, one grating or a turret with 4 gratings from the list below*

Grooves/mm1) 2400 1800 1200 600 400 300
Reciprocal linear dispersion (nm/mm) 1.58 2.12 3.18 6.37 6.32 9.6 9.41 12.75 12.64
Blaze wavelength (nm) 225 270 400 750 1000 800 1700 1500 2000
Spectral range (nm) 190–450 190–540 265–800 500–1500 660–800 530–1600 1130–2600 1000–3000 1330–3600
Multi-channel array bandpass (average), nm 384) 514) 764) 1504) 805) 2304) 1205) 1605) 1605)
Spectral resolution (nm) <0.14) <0.154) <0.224) <0.454) <0.455) <0.74) <0.75) <0.945) <0.945)
Wavelength accuracy (nm) ± 0.15 ± 0.217 ± 0.32 ± 0.65 ± 1.00 ± 1.34
Wavelength repeatability (nm) ± 0.113 ± 0.167 ± 0.245 ± 0.50 ± 0.775 ± 1.04
Step of grating rotation (nm) 0.007 0.009 0.013 0.026 0.04 0.05
Scanning rate (nm/s) 12.5 16.7 25 50 75 100
  • 1) Diffraction grating with an alternate number of grooves per mm/blaze wavelength can be used upon your request.
  • 2) Reciprocal linear dispersion is indicated for blaze wavelength.
  • 3) Wavelength range with diffraction efficiency higher than 40%.
  • 4) For the detector with 24 µm pixel size and 24.5 mm of active area.
  • 5) For the detector with 25 µm pixel size and 12.8 mm of active area.